Tri-axial Fluxgate Probe
Tri-axial Fluxgate Probe
Number of axsis |
3(XYZ coordinate on right) |
Polarity |
Non-inverting output when +ve points north |
Measurement range |
±70μT (tickler coil current=±1mA ) |
- 3dB Bandwidth |
> 100 Hz |
Noise (normal) |
<300pTrms /√Hz 1Hz full ℃range |
Noise (low) |
<30pTrms /√Hz 1Hz full ℃range |
Standard Error |
±5% (25°C), ±10% (215°C) |
Compensation |
±500nT (max),25°C, ±1000nT 215°C |
Start/stabilization time |
0.5s, about 99% of the end value |
orthogonal error between axis |
< 3° (max) |
linear error |
0.005% (least square fit) |
frequency response |
< 1%of amplitude, DCis about 10 Hz |
hysteresis |
< 100nT Exposed in 1mT Environment |
Condition
Operatingtemperature range |
0°C to +215°C |
Storage temperature |
-55°C to +215°C |
protect/package; pressure |
IP68, 3 bar |
Thermal shock, shake and other pressure |
1.3ms 150G (max) |
Machinery Index
Size |
23mm3 |
Weight |
30g |
Fitting |